Description; Figure 2 Design Of The Conditioning And Sampling System Ss-5 And Ss-5/3 - M&C PSS SS-5 Instruction Manual

Gas conditioning unit
Table of Contents

Advertisement

9

DESCRIPTION

Powe
Status contact
r
suppl
y
385
C ON
C
Figure 2
Design of the conditioning and sampling system SS-5 and SS-5/3
The installation of a gas cooler  and an appropriate diaphragm gas pump  will be carried out
according to the maximum flow of gas extracted (see instruction manual for individual components).
The possible combinations are listed in the following table:
SS-5...
Type of
cooler
SS-5
ECP1000
SS-5/3
ECP3000
The minimum amount of flow is determined by the sample gas pump (see 8.). Premature damage can
be caused to the pump membrane if less than the minimal total amount of flow is extracted as a result
of excess pressure.
All gas coolers are equipped with a Duran glass heat exchanger . Heat exchangers of PVDF or
stainless steel are also available.
www.mc-techgroup.com
515
Option
OUT
LA1.2
TRM1.2
GL18
Cooler
EC
P
GL25
max. gas flow [Nl/h]
150
350
SS-5, SS-5/3 | 1.00.01
1 = Sample gas IN
2 = Sample gas OUT
3 = Condensate OUT
(all connections DN4/6)
1
2
3
Sample gas pump
N 3 KPE
N 9 KPE
9