Guidance And Manufacturer's Declaration - Electromagnetic Immunity - Care Fusion Alaris Technical & Service Manual

Gateway workstation
Table of Contents

Advertisement

Guidance and Manufacturer's Declaration - Electromagnetic Immunity

The Workstation is intended for use in the electromagnetic environment specified below .
The customer or the user of Workstation should assure that it is used in such an environment .
Immunity Test
EN 61000-4-2 Electro-Static
Discharge (ESD)
EN 61000-4-4
Electrical Fast Transient, Burst
(EFT) (Note 3)
EN 61000-4-5
Power Line Surge
(Note 3)
EN 61000-4-8 Power Frequency
Magnetic Field (50/60 Hz)
EN 61000-4-11
Voltage Dips, Short
Interruptions, and Voltage
Variations
(Note 3)
Note 1—UT is the AC mains voltage prior to application of the test level .
Note 2—Compliance levels raised by EN 60601-2-24 .
Note 3—Performed at the Minimum and Maximum Rated Input Voltage .
Note 4—CareFusion recommends using signal cables of less than 3 metres in length and this requirement is applicable only if signal
cables are 3 metres or more in length . (EN 60601-1-2:2002, Clause 36 .202 .4)
EN 60601-1-2
Compliance Level
Test Level
±6 kV contact
±8 kV contact (Note 2)
±8 kV air
±15 kV air (Note 2)
±2 kV for power
±2 kV for power supply
supply lines
lines
±1 kV for input/
N/A (Note 4)
output lines
±1 kV Line(s) to
Line(s)
±1 kV Line(s) to Line(s)
±2 kV Line(s) to
±2 kV Line(s) to Earth
Earth
3 A/m
400 A/m 50 Hz (Note 2)
<5 % UT (Note 1)
<5 % UT
(>95 % dip in UT)
(>95 % dip in UT)
for 0 .5 cycle
for 0 .5 cycle
40 % UT
40 % UT
(60 % dip in UT)
(60 % dip in UT)
for 5 cycles
for 5 cycles
70 % UT
70 % UT
(30 % dip in UT)
(30 % dip in UT)
for 25 cycles
for 25 cycles
<5 % UT
<5 % UT
(>95 % dip in UT)
(>95 % dip in UT)
for 5 sec
for 5 sec
1000SM00015 Issue 8
Alaris® Gateway Workstation
Electromagnetic Environment – Guidance
Floors should be wood, concrete, or ceramic tile .
If floors are covered with synthetic material, the
relative humidity should be at least 30 % .
Mains power quality should be that of a typical
commercial or hospital environment .
Mains power quality should be that of a typical
commercial or hospital environment .
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment .
Mains power quality should be that of a typical
commercial or hospital environment .
If the user of the Workstation requires continued
operation during power mains interruptions,
it is recommended that the Workstation be
powered from an uninterruptible power supply
or a battery .
The Workstation does employ an internal short
duration battery .
59/72