Mounting Locations
It is important that the mounting brackets are located exactly as detailed in the diagrams below . Each mounting bracket must be
attached to a load bearing Dräger type mounting rail when the Bar Mounting kit is used or a load bearing pole when the Pole Mounting
kit is used .
The Workstation should not be fitted to any other mobile pole or drip stand unless the stability and strength of the
A
whole assembly has been evaluated to IEC/EN60601-1 .
Note: The mounting
brackets detailed in the
following diagrams are the Bar
Mounting Kit (1000SP00655) .
The indicated mounting
positions also apply to the Pole
Mounting Kit (1000SP00169) .
3x0, 3x2, 3x3 and 3x4 Configuration
Mounting kits required: 2
1000SM00015 Issue 8
Alaris® Gateway Workstation
53/72
Mounting Options