IEC Pod/Rear Cover

Replacement Procedure
1 . Ensure the Mains Inlet Cable Assembly
'Power Supply Unit')
2 . Remove the eight screws
3 . Remove the PE Connector Nut
4 . Remove the Ethernet Earth Cable, the Services Earth Cable and the Mains Inlet Cable crimp connectors .
5 . Remove the Shakeproof Washer
K
6 . Remove the PE Stud
and Earth Symbol Washer
7 . Remove the three Screws
8 . Remove the Mains Inlet Cable Assembly
9 . Reassemble in reverse order .
Refitting Note:
H
E
Torque Nuts
and
to 150cNm . Torque Screw
The Fuses are located in the Mains Inlet Cable Assembly and can be replaced without disassembling the Workstation .
A
Replace the Fuses as required .
B
(x8)
O
Seal
N
IEC Pod Cover
M
(x3)
L
(x3)
A
is disconnected from the Tile PCB (see '3 Tile Assembly') and the Power Supply Unit (see
B
C
and washers
securing the Rear Cover
E
F
, Washer
.
G
H
, Nut
and M6 Washer
J
.
L
M
and Washers
securing the IEC Pod Cover
A
from the slot in the Rear Cover Assembly .
B
and
D
Rear Cover
C
(x8)
A
Mains Inlet
Cable Assembly
K
PE Stud
J
Earth Symbol
Washer
1000SM00015 Issue 8
D
to the Power Base Module .
I
.
N
and Seal
L
to 45 cNm .
Alaris® Gateway Workstation
Corrective Maintenance
O
. Remove the IEC Pod Cover .
I
Washer
H
Nut
G
Washer
F
Washer
E
Nut
32/72