Tiles not Shown in the Web User Interface or PDMS (Software v1.1.3)

Software configuration for the tile's Ethernet Switches was corrected in v1 .1 .5 and above .
• Upgrade to the latest Workstation software variant .

Tile Failure and Portal System Fault Indicator (SFI) alarms when Pump is fitted to a Tile

• Replace the Tile assembly and upgrade the Workstation to the latest software variant
• Send Customer Feedback and faulty Tile assembly to the Customer Advocacy department at CareFusion if the following revision
status is shown on the tile PCB:
-
Two Vertical high - 1000EL00708 Rev 8 and above
-
Two Horizontal wide - 1000EL00712 Rev 8 and above
-
Three Vertical high - 1000EL00636 Rev 10 and above
-
Three Horizontal wide - 1000EL00716 Rev 7 and above

System Fault Indicator (SFI) Alarms - Compact Flash (CF) Card Failures

The Workstation alarms 5 minutes after every power up .
There is no communication with the Workstation Web Service page or the event log failed to download .
• It is recommended that all Compact Flash Cards are purchased directly from CareFusion, as the use of Compact Flash Cards
purchased from any other source may invalidate the warranty . Please contact CareFusion Technical Support for further guidance
• Replace the CF card
Note: The 128mb
Compact Flash Card
has now been replaced
by the larger capacity
512mb Compact Flash
Card .
1000SM00015 Issue 8
Alaris® Gateway Workstation
17/72
Troubleshooting