7 Appendix

Electromagnetic Compatibility

Warning

• The use of any accessory, transducer, or cable with the Workstation other than those specified may result in increased emissions
or decreased immunity of the Workstation .
• The Workstation should not be used adjacent to or stacked with other equipment, however if adjacent or stacked use is necessary,
the Workstation should be observed to verify normal operation in the configuration in which it will be used .

Caution

• The Workstation is a CISPR 11 Group 1 Class A Medical Equipment System and intended for use by healthcare professionals only .
• Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed, put into service and used
according to the EMC information provided in the accompanying documents .
• Portable and Mobile RF communications can affect Medical Electrical Equipment .
• Operating the Workstation near equipment which radiates high energy radio frequencies (electro surgical or cauterizing
equipment, portable radios, cellular telephones, etc .) may cause false alarm conditions . If this happens, reposition the
Workstation away from the source of interference or turn off the Workstation and manually regulate the flow .
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The Workstation is intended for use in the electromagnetic environment specified below .
The customer or the user of the Workstation should assure that it is used in such an environment .
Emissions Test
CISPR 11
RF Emissions
CISPR 11
RF Emissions
EN 61000-3-2
Harmonic Emissions
EN 61000-3-3
Voltage Fluctuations,
Flicker Emissions
Compliance
Electromagnetic Environment – Guidance
The Workstation uses RF energy only for its internal function in the normal
Group 1
product offering . Therefore, its RF emissions are very low and are not likely to
cause any interface in nearby electronic equipment .
Class A system when
used in conjunction
with pumps
Class B in stand alone
operation
The Workstation is suitable for use in all establishments, other than domestic,
and those directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes .
Class A
Complies
1000SM00015 Issue 8