Battery Test
Rear View of
Auxiliary Workstation
This test verifies that the battery is functioning correctly, with and without AC mains attached to the Workstation:
Barcode Reader
Connector
Nurse Call
Connector
Rear View of
Workstation
1 . Connect the Workstation to the test equipment as shown . Connect Workstation to AC mains using IEC Mains Lead .
2 . Connect the Cat5e crossover cable between the Auxiliary Connector port on the rear of the Workstation and the test PC ethernet
port .
3 . Switch on the Workstation . Wait for the Workstation to initialise (approx . 45 seconds) .
4 . Manually set the test PC's IP address to 192 .168 .0 .130, Subnet mask 255 .255 .255 .0 .
5 . On the Test PC, launch Internet Explorer . In the address bar at the top of the screen enter the following address: http://192.168.0.128
6 . Confirm that the Workstation web service startup page is displayed . Click the Refresh icon and confirm that the startup page
continues to be displayed when the refresh is complete .
7 . Click OK on the startup screen .
8 . Check that, at the bottom right hand corner of the monitoring page on the test PC, the battery symbol shows a capacity percentage
of between 5% and 100% . If the capacity is less than 5% then the Workstation must remain connected to AC mains until the capacity
reaches 50% .
9 . Remove the AC mains lead and ensure that the Workstation continues to operate and shows a capacity percentage of between
5% and 100% . If the unit does not operate or shows a percentage less than 5% then it is likely that a battery sub-system fault has
occurred and should be replaced only after further investigation .
Rear View of
Workstation
Test Equipment
Cat5e Ethernet
Base Module
Crossover Cable
(1000JG00275)
Cat5e Ethernet
Test PC
Crossover Cable
with Wireless LAN card
1000LB00644 Iss 2
1000LB00644 Iss 2
1000LB00644 Iss 2
Test Equipment
Base Module
(1000JG00277)
Cat5e Ethernet
Test PC
Crossover Cable
with Wireless LAN card
1000SM00015 Issue 8
Alaris® Gateway Workstation
Test PC
12/72
Routine Maintenance