Setting the Time and Date

1 . Ensure that the Cat5e crossover cable between the Auxiliary Connector port on the rear of the Workstation and the test PC ethernet
port .
2 . Connect the Workstation to AC mains using an IEC Mains Lead . Switch on the Workstation .
3 . Manually set the test PC's IP address to 192 .168 .0 .130, Subnet mask 255 .255 .255 .0 .
4 . On the Test PC, launch Internet Explorer . In the address bar at the top of the screen enter the following address: http://192.168.0.128
5 . Click OK on the startup screen . Confirm that the tile monitoring page is displayed .
6 . Click on Date and Time in the navigation panel on the left of the screen . The following screen is displayed:
This screen shot is for illustrative purposes only and may not reflect the installed software version.
7 . Enter the correct date and correct time .
8 . Click Update to confirm .
9 . Click on Event Logs in the navigation panel on the left of the screen . An entry for Time Changed and Date Changed at the bottom
of the list indicates that the time and date have been updated successfully .