PANNORAMIC® SCAN RX 1.1
Warning!
Only connect external devices that are safety extra low voltage rated to the instrument to
avoid the risk of electrical shock.

2.4 Coupling the Lumencor SOLA U-NIR unit with the scanner

To operate the Lumencor SOLA U-NIR fluorescent illumination unit, it has to be connected
to the Pannoramic® SCAN RX base unit.
1.
Turn the control computer and the Pannoramic® SCAN RX unit off.
2.
Connect both ends of the fiber optic cable to the coupling ports of the scanner and the
illumination unit (see Figure 2 – option A in chapter
3.
Connect the power supply cables of both the scanner and the illumination unit.

2.5 Coupling the Lumencor Spectra 6 unit with the scanner

To operate the Lumencor Spectra 6 Light Engine fluorescent illumination unit, it has to be
connected to the Pannoramic® SCAN RX base unit, the camera, and the control computer.
1.
Turn the control computer and the pco.edge 4.2 bi camera off.
2.
Attach Lumencor SPECTRA 6 fluorescent illumination unit to the coupling adapter of the
Pannoramic® SCAN RX scanner (see Figure 2 – option B in chapter
3.
Connect power supply cables of both the scanner and the illumination unit.
4.
Connect TTL trigger cable between pco.edge 4.2 bi camera (out 4) and the Lumencor
SPECTRA 6 unit (TTL port), and USB-to-RS232 communication cable between the
Lumencor SPECTRA 6 unit and the computer.
NOVEMBER 18, 2021
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
Figure 5. Lumencor SPECTRA 6 Light Engine connecting ports
1.1.4 Base
Unit).
1.1.4 Base
INSTALLATION
Unit).
28