PANNORAMIC® SCAN RX 1.1
Click the
button to display barcode label area. To rotate the label area by 180°, click
.
Undo / Redo modification made during specifying the Scanmap. These buttons
are inactive when Use threshold function is applied.
5.1.2.1
Focus indicator
Click
to activate the Focus indicator function. Search for the sharpest focus on the
current field of view automatically by clicking
To define focus limits, select Limit the range. Set the upper and lower limits by selecting the
desired value in the selection boxes. By using ± Range, the Lower limit value is set to -32 of
the current focus value and the Upper limit value is set to +32 of the current focus value.
This is useful if Auto Focus does not find the focus on the specimen, but on the coverslip.
Also, this makes the focusing sequence faster.
For manual focusing define the manual focus step size by adjusting the Step size slider and
use the two arrow buttons to focus manually. This is useful when you are interested in an
area that remained blurry with Auto Focus (for example, the specimen is thicker in that
area). This does not influence the way slides are digitized in Auto focus mode.
NOVEMBER 18, 2021
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
Figure 51: Barcode label area window
Figure 52. Focus indicator panel
SCANNING MODES AND SETTINGS
Auto focus button.
75