3DHISTECH PANNORAMIC SCAN RX 1.1 User Manual page 157

Table of Contents

Advertisement

PANNORAMIC® SCAN RX 1.1
1.
Turn off the fluorescent illumination unit.
Warning!
Crush hazard! Stop the running digitization process and turn off the scanner (refer to
chapter
2.7 Power the Scanner
removal.
2.
Select the desired filter in the corresponding software window (see chapter
The filter will be positioned in such a way that the user can access it without having to
turn the filter wheel manually.
3.
Power off the scanner. See chapter
device.
4.
Remove the protective cover of the device (see chapter
Covers).
5.
Hold the filter module (3) by its handling pads (one on the right, one on the left (lateral)
side of the module) and insert it in forward tilted position from above into the upper
spring-loaded clamps (1) at the filter wheel.
Warning!
Avoid touching the surface of the filters and store the modules in their own boxes when not
used.
6.
Press the filter module down, until it safely connects to the bottom spring-loaded clamps
(2) of the filter wheel. Mounting position of the filters: excitation filter (5); emission filter
(4).
7.
To remove the filter module, first slightly tilt it out of its upper spring-loaded fixtures, then
pull it out of the bottom spring-loaded components (1) and take it off.
8.
Place the protective cover back again (see chapter
9.
Power on the scanner. See chapter
device.
10.
Enter the respective parameters in the input fields e (see chapter
11.
Turn the fluorescent illumination unit on again (consider the cooling time possibly
required).
NOVEMBER 18, 2021
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
On/Off) before you proceed with filter module installation or
2.7 Power the Scanner On/Off
2.7 Power the Scanner On/Off
TROUBLESHOOTING AND MAINTENANCE
on how to power off the
9.5.1 Removing the Protective
9.5.1 Removing the Protective
on how to power on the
6.1
Settings).
6.1
Settings).
Covers).
156