3DHISTECH PANNORAMIC SCAN RX 1.1 User Manual page 91

Table of Contents

Advertisement

PANNORAMIC® SCAN RX 1.1
Preview Types
In FL Microscope mode the following additional preview modes are available:
o
Brightfield preview
is recommended when scanning in Brightfield mode.
o
Pre-scan
Wizard window. Scanning process takes a couple of minutes only, and it cannot be
interrupted.
o
Pre-scan high-resolution
of every FOV within the selected area (in single layer DAPI, with an illumination
time of 20ms) that is displayed in the Profile Wizard window. Scanning process
takes a couple of minutes only and cannot be interrupted.
o
Pre-scan + high-resolution
image of every FOV within the selected area (in single layer DAPI, with an
illumination time of 20ms) that is displayed in the Profile Wizard window.
Scanning process is initiated with a simple Pre-scan, the located sample with auto
threshold is selected, and finally, a Pre-scan in high-resolution is launched.
Remove coverslip
If selected, the edges of the coverslip are not detected as specimen to be scanned.
Depending on the threshold level, it is possible that the edges of the coverslip are detected
as areas to be digitized.
NOVEMBER 18, 2021
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
– a normal type of illumination for scanning. This option
– creates a fast FL preview image that is displayed in the Profile
– creates a High-Resolution fast FL preview image
– creates a High-Resolution fast FL preview
Figure 72. Preview image with and without coverslip edge
SCANNING MODES AND SETTINGS