PANNORAMIC® SCAN RX 1.1
Basic settings
o
On this panel, you can define the swap file location in the Swap location field, set
the value of Shutter auto close time in FL, and define a value in the Number of
compensation images the preview camera should take, if needed.
o
By activating the Inspect area beyond coverslip edge function, a compensation
image of the area beyond the coverslip area can be taken.
Consequently, when this function is turned off, the area beyond the coverslip is
not taken into account by the scanner software when creating the compensation
image.
o
For a more comfortable and safer use, turn on Planner Autosave function to save
a scan plan, and by defining a location in the Planner autosave folder field the
scan plan will be automatically saved.
o
Set the character as value separator used in the Planner export file at the CSV
separator field.
o
Click Apply to accept modifications.
NOVEMBER 18, 2021
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
Figure 112. Basic settings