PANNORAMIC® SCAN RX 1.1
o
Use marker with threshold
If the specimen cannot be identified as required for digitization, (for example,
weak contrast), it can help to encircle completely the specimen with a black
marker pen. You have to draw a full circle with the pen to enable the detection.
Marker pen detection will make specimen detection easier by limiting the scanned
area.
If this option is selected, only the area encircled with a marker pen will be
scanned. You can define the threshold level by changing the threshold value in the
Use marker with threshold selection box in the range of 0-255. This sets the
sensitivity of the specimen detection inside the marker area.
To set the desired threshold value manually, change the setting in the Specimen
threshold selection box in the range of 0-255. The default value is 105.
o
Everything within marker – To scan everything within the marked area, check the
Everything within marker selection box.
Alternatively, click
Figure 37. Scan everything within the marked area (selected / not selected)
NOVEMBER 18, 2021
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
in the preview window to activate this function.
SCANNING MODES AND SETTINGS
60