Example 9: Pass/Fall Test

The Pass/Fail Test is one of the enhanced special functions based on VS5000 series.
By this function, the oscilloscope could compare the input signal with the established
waveform mask. If the waveform "touches" the mask, a "Fail" signal occurs.
Otherwise the test passes. When needed, a programmable output can be used for
external automatic control applications. The output is built in as a standard feature
and is optically isolated.
Do the steps as follows:
1.
Click Utility
2.
Select the Source to CH2 (the input source is CH2). Output to PASS+BEEPER.
Select the
option.
3.
Create the Pass/Fail Mask: Set the div of the Vertical and Horizontal value.
Click Save or Load to save or load the P/F Mask.
4.
Click the Run button in the Operation window to start the Pass/Fail function.
3-12
Pass/Fail to show the Pass/Fail Setup window.
option to
in front of the Stop When Pass/Fail to choose this
User's Guide for VS5000 Series