Horizontal
Sample Rate Range
Waveform
interpolation
Memory Depth
Time Base Range
(sec/div)
Sample Rate and
Delay Time Accuracy
Delta Time
Measurement
Accuracy
(Full Bandwidth)
5-2
1Sa/s-400MSa/s (Real-Time sample), 25Sa/s (Equivalent
sample)
Sin(x)/x
1M samples for single channel, 512k samples for each
channel
[4]
512K samples
2ns/div-50s/div, VS5202X, VS5102X
5ns/div-50s/div, VS5062X
10ns/div-50s/div, VS5042X
20ns/div-50s/div, VS5022X
1-2-5 Sequence
±50ppm (over any 1ms time interval)
Single: ±(1 sample interval + 50ppm × reading + 0.6 ns)
>16 averages: ±(1sample interval + 50ppm × reading + 0.4
ns)
User's Guide for VS5000 Series