Installation (Electrical) - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Compact controller measured variable: ph / orp
Table of Contents

Advertisement

5.3.3

Installation (electrical)

The cable must be routed in a site-
provided cable duct to ensure
strain relief
1.
Undo the four housing screws
2.
Slightly lift the controller housing
top section forwards and fold it to
the left
Fig. 16: Punch out threaded holes
3.
Large threaded connection
(M 20 x 1.5)
Small threaded connection
(M 16 x 1.5)
Punch out as many threaded con‐
nections on the bottom side of the
controller housing bottom section
as required
4.
Guide the cable into the respective
reducing inserts.
5.
Insert the reducing inserts into the
threaded connectors
6.
Guide the cable into the controller.
7.
Connect the cable as indicated in
the terminal diagram
8.
Screw the required threaded con‐
A0272
nections in and tighten
9.
Tighten the clamping nuts of the
threaded connections so that they
are properly sealed
10.
Click the controller housing top sec‐
tion on to the controller housing
bottom section
11.
Manually tighten the housing
screws
12.
Once again check the seating of the
seal. Only if the mounting is correct,
is protection class IP 67 (wall/pipe
mounting) or IP 54 (control panel
mounting) achieved
Assembly and installation
39