Maintenance; Changing The Fuse, Dulcometer ® Compact Controller - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Compact controller measured variable: ph / orp
Table of Contents

Advertisement

10

Maintenance

Users' qualification: trained user, see
n
Ä Chapter 2.2 'Users' qualifications'
on page 10
The DULCOMETER
is maintenance free.
10.1
Changing the fuse,
DULCOMETER
pact Controller
WARNING!
Danger from electrical voltage
Possible consequence: Fatal or
very serious injuries.
The DULCOMETER
Controller does not have a
mains switch
When working inside the con‐
trol unit, disconnect the control
unit from the mains power via
an external switch or by
removing the external fuse
NOTICE!
Use only 5 x 20 mm micro-fuses
Possible consequence: Damage to
the product or its surroundings
5x20 T 0.315 A
Part number 732404
Compact Controller
®
Com‐
®
Compact
®
Fuse change
The mains fuse is located in a sealed fuse
holder in the inside of the device.
1.
Disconnect the controller from the
mains power
2.
Open the controller and fold the
controller housing top section to the
left
3.
Remove the PCB cover
4.
Remove the micro-fuse using a
suitable tool
5.
Fit the micro-fuse using a suitable
tool
6.
Refit the PCB cover
7.
Replace controller housing top sec‐
tion and close the controller
Maintenance
83