Operating menus for the measured variables pH and ORP
During calibration, the DULCOMETER
'0' . Exception to this: a basic load or a manual control variable has been set. This
remains active. The mA standard signal output is frozen.
When calibration/testing has been completed successfully, all of the error checks relating
to the reading are restarted. The DULCOMETER
determined data for zero point and slope when the calibration is successful.
Used buffer
Dispose of the used buffer solution. Related info: see buffer solution safety data
sheet.
Setting
Starting
value
Buffer tem‐
Measured
perature
value
Buffer
Start value =
values
7.00 pH
(ZERO)
4.00 pH
(SLOPE)
50
Compact Controller sets the control outputs to
®
®
Possible values
Increment
Lower value
0.1 ℃
0 ℃
0.01 pH
0.00 pH
Compact Controller saves all the
Upper value
120 °C
14.00 pH
Remarks
The tem‐
perature can
only be
adjusted
under
'TEMP'
'auto'
or
'manual'
Limit value
ZERO = 6..8
pH
Limit value
SLOPE = <
6 pH; > 8 pH