2.
Fig. 50: Dismantling the hinge
3.
Remove the screw (2), unclip the
hinge (1) on the lower part of the
controller housing (arrows) and
remove the hinge
1.
1.
2.
1.
Replacing spare part units
Fig. 51: Removing the strain relief
4.
Remove the strain relief (2).
Remove the screws (1) to do so.
5.
Check the profile seal (arrow), then
position the profile seal evenly into
the groove in the top part of the
DULCOMETER
troller housing. Arrange the flaps (3)
as shown in the figure
6.
Remove the top part of the con‐
troller housing (3 fixing bolts)
7.
Label the cable connectors fitted to
distinguish them and remove the
cables from the lower part of the
controller
Preparing the new lower part of the con‐
troller housing
A0351
Fig. 52: Punching out the threaded holes
8.
Large threaded connection
(M 20 x 1.5)
Small threaded connection
(M 16 x 1.5)
Punch out as many threaded holes
on the bottom of the lower part of
the controller housing as required
®
Compact Con‐
A0272
99