11.16 Group CALIBRATION DATA

Functional description CALIBRATION DATA
CAL. FACTOR
Use this function to call up the calibration factor currently used.
Data indicated:
5-digit floating-point number (typically 1.000)
ZERO POINT
Use this function to call up the zero point correction currently used.
Data indicated:
5-digit floating-point number with sign (e.g. +0200.0)
ZEROPOINT STAT.
Use this function to call up or manually change the static zero point correction currently
being used.
User input:
5-digit floating-point number, including unit and sign (e.g. +0010.0 ns)
CORR. FACTOR
Use this function to enter a correction factor at the client's site.
User input:
5-digit floating-point number between 0.5 and 2.
Factory setting:
1.000 (no correction)
Description of device functions
95