Cable Specifications - Endress+Hauser Proline Prosonic Flow 91 Operating Instructions Manual

Ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 91:
Table of Contents

Advertisement

Proline Prosonic Flow 91
Endress+Hauser
Fig. 23:
Connecting nameplate for sensor connecting cables
4.1.2

Cable specifications

Sensor cable
• Use the ready-to-use cables supplied by Endress+Hauser with each sensor pair.
• The following cable lengths are available:
– 5 m, 10 m, 15 m , 30 m, 60 m
– 16 ft, 33 ft, 49 ft, 98 ft, 197 ft
• Cable material: PVC
• Operating temperature: –20 to +70 °C (–4 to +158 °F)
Operation in zones of severe electrical interference:
The measuring device complies with the general safety requirements in accordance with
EN 61010-1, the EMC requirements of IEC/EN 61326 "Emission to class A requirements", and
NAMUR Recommendation NE 21.
Wiring
a0005875