Commissioning
!
40
6.3

Commissioning via a configuration program

Sensor Setup/sensor installation

6.3.1
For sensor installation with the configuration program "FieldCare" any "Quick Setup" menus which
correspond to the local operation exist.
Other methods (see table) are available for determining the relevant values for sensor distance, wire
length, etc. The procedure is illustrated in detail on → ä 40.
Sensor type
Required values for the sensor
installation procedure
clamp-on
Position sensor
version
Wire length
Sensor distance
1)
Conditions that must be met before determining the values via the local display using the
"Sensor" Sensor Setup (→ ä 70):
• Transmitter installed (→ ä 21)
• Transmitter connected to power supply (→ ä 24)
2)
FieldCare is a configuration software package for flowmeters in the field. Conditions that must be met before
determining the values via "FieldCare":
• Transmitter installed (→ ä 21)
• Transmitter connected to power supply (→ ä 24)
• "FieldCare" configuration package installed on a notebook/PC
• Connection made between notebook/PC and device via the FXA291 service interface (→ ä 24)
3)
Applicator is software for selecting and configuring flowmeters. The values required can be determined without having
to connect the transmitter beforehand.
The "Applicator" can be downloaded from the Internet (→ www.applicator.com) or ordered on
CD-ROM for installation on a local PC.
Procedure (determining data for sensor installation)
The following table can be used to select and configure, the functions required to install the sensor:
Note!
Enter a valid release code to allow that device parameters can be changed or activated. The code
(factory setting = 91) is entered by means of the corresponding matrix cell.
"Clamp-on" sensor installation
Procedure
Selection - Input - display
Liquid in the pipe
Liquid temperature
Liquid sound velocity
Pipe material
Pipe sound velocity
Pipe circumference
Pipe diameter
Wall thickness
Liner material
Liner sound velocity
Liner thickness
Sensor type
Sensor configuration
1)
Local display
x
x
x
Local display (Sensor Setup)
LIQUID
TEMPERATURE
SOUND VELOCITY LIQUID
PIPE MATERIAL
SOUND VELOCITY PIPE
CIRCUMFERENCE
PIPE DIAMETER
WALL THICKNESS
LINER MATERIAL
SOUND VELOCITY LINER
LINER THICKNESS
SENSOR TYPE
SENSOR CONFIGURATION
Proline Prosonic Flow 91
2)
3)
FieldCare
Applicator
x
x
x
x
x
x
Endress+Hauser