Installation For Measurement; Via One Traverse - Endress+Hauser Proline Prosonic Flow 91 Operating Instructions Manual

Ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 91:
Table of Contents

Advertisement

Proline Prosonic Flow 91
Endress+Hauser
4.
Interlock both halves of the lock (3). Make sure that there is sufficient space for the tensioning
band to be tightened with the locking screw.
5.
Tighten the tensioning band using a screwdriver (4).
1
2
3
Fig. 10:
Tensioning band installation for DN 250 to 4000 (10" to 160")
3.3.2
Installation for measurement via one traverse
Prerequisites
• The installation distances (sensor distance and wire length) are known → ä 12.
• The strapping bands are already mounted → ä 14.
Material
The following material is needed for mounting:
• Two strapping bands incl. mounting bolts and centering plates where necessary
(already mounted → ä 14)
• Two measuring wires, each with a cable lug and a fixer to position the strapping bands
• Two sensor holders
• Coupling fluid for an acoustic connection between the sensor and pipe
• Two sensors incl. connecting cables.
Procedure
1.
Prepare the two measuring wires:
– Arrange the cable lugs and fixer such that the distance they are apart corresponds to the wire
length (SL).
– Screw the fixer onto the measuring wire.
Fig. 11:
Fixer (a) and cable lugs (b) at a distance that corresponds to the wire length (SL)
SL
1
2
3
4
5
6
7
8
Installation
9
10
11
a0001110
A0001112
15

Advertisement

Table of Contents
loading

Table of Contents