Description of device functions
90

11.14 Group LIQUID DATA

LIQUID
Use this function to select the liquid in the pipe.
Options:
WATER
SEA WATER
DISTILLED WATER
AMMONIA
ALCOHOL
BENZENE
BROMIDE
ETHANOL
GLYCOL
KEROSENE
MILK
METHANOL
TOLUOL
LUBE OIL
FUEL OIL
PETROL
OTHER
Factory setting:
WATER
!
The selection specifies the values for the sound velocity and viscosity. If OTHER is
selected, these must be entered via the SOUND VELOCITY LIQUID and VISCOSITY
functions.
TEMPERATURE
Use this function to enter the process temperature of the liquid. Via the sound velocity,
the value influences the determination of the sensor distance. Enter the process
temperature at normal operating conditions to achieve an optimum configuration of the
measuring system.
User input:
Fixed-point number –273.15 °C to 726.85 °C
Factory setting:
20 °C
Functional description LIQUID DATA
Note!