Group User Interface - Endress+Hauser Proline Prosonic Flow 91 Operating Instructions Manual

Ultrasonic flow measuring system
Hide thumbs Also See for Proline Prosonic Flow 91:
Table of Contents

Advertisement

Proline Prosonic Flow 91
Endress+Hauser
11.6

Group USER INTERFACE

Functional description USER INTERFACE
FORMAT
Use this function to define the maximum number of places after the decimal point
displayed for the reading in the main line.
Options:
XXXXX.
XXXX.X
XXX.XX
XX.XXX
X.XXXX
Factory setting:
X.XXXX
!
• Note that this setting only affects the reading as it appears on the display, it has no
• The places after the decimal point as computed by the measuring device cannot always
CONTRAST LCD
Use this function to optimize display contrast to suit local operating conditions.
User input:
10 to 100%
Factory setting:
50%
TEST DISPLAY
Use this function to test the operability of the local display and its pixels.
Options:
OFF
ON
Factory setting:
OFF
Test sequence:
1. Start the test by selecting ON.
2. All pixels of the main line and additional line are darkened for at least 0.75 seconds.
3. The main line and additional line show an "8" in each field for at least 0.75 seconds.
4. The main line and additional line show a "0" in each field for at least 0.75 seconds.
5. The main line and additional line show nothing (blank display) for at least
When the test completes the local display returns to its initial state and the setting changes
to "OFF".
Note!
influence on the accuracy of the system's calculations.
be displayed, depending on this setting and the engineering unit. In such instances an
arrow appears on the display between the measuring value and the engineering unit
(e.g. 1.2 → l/h), indicating that the measuring system is computing with more decimal
places than can be shown on the display.
0.75 seconds.
Description of device functions