14 Decommissioning

Safety instructions

CAUTION
Risk of injury due to corrosive condensates
Condensates are often corrosive.
When working with corrosive reagents note the hazard
information and safety precautions contained in the
applicable material safety data sheets.
Neutralize condensates and follow the prescribed
measures for disposal.
NOTICE
Damage to components!
Damage to components due to improper decommissioning.
Before being shut down the analyzer system should be
purged in order to prevent condensation and condensate
deposits in the individual units.

Shutting Down the Analyzer System

1. Flush the sampling probe, filter and sample gas line, e. g. by
drawing outside air from the sampling probe.
2. Purge the gas paths of the analyzer system for 30 minutes.
3. Turn off the analyzer system with main switch –Q60.
Emptying the Condensate Collecting
Bottle
Empty the condensate collecting bottle and dispose of
condensates according to applicable regulations.
Note
Make sure the analyzer system is free of residual moisture that
can freeze if low temperatures are encountered during shipping
and storage.

Transportation restraints activation

1 Sample Gas Feed Unit SCC-F transportation restraints
2 Sample Gas Cooler SCC-C transportation restraints
Figure 61: Position of transportation restraints (Bottom view of analyzer cabinet)

Sample Gas Feed Unit SCC-F

Diaphragm Pumps transportation restraints:
Using a Ph2 crosshead screwdriver, screw two M6x25 screws
through the holes
1
in the base plate into the diaphragm
pumps base plate and tighten them.

Sample Gas Cooler SCC-C

Compressor transportation restraints:
Using an offset Ph2 crosshead screwdriver, turn the two
screws clockwise through the holes
the point at which the compressor housing is in contact with
the base plate (noticeable resistance).

Ambient conditions

Ambient temperature during transport / storage
2 to 60 °C (35.6 to 140 °F);
−20 to 70 °C (−4 to 158 °F) after draining and drying parts
in contact with condensate.
Max. permissible humidity
Year-round average max. 75%, short-term max. 95%,
occasional slight condensation is permitted.