Installation - ABB CEMcaptain GAA610-M Operating Instruction

Advanced emission gas monitoring system for marine applications
Table of Contents

Advertisement

26
GAA610-M ADVANCED EMISSION GAS MONITORING SYSTEM | OI/GAA610-M-EN REV. B

... 6 Installation

... Probe tube and filter unit installation
Installation
Flange orientation
1 Duct wall
2 Flange
3 Flange gasket
4 Probe flange
Figure 12: Filter unit installation
1. Screw the probe tube
6
2. Insert the pre-assembled probe tube with filter unit in the wall tube and screw the flange of the filter unit to the flange of the wall
tube. Use the green seal from the accessories pack to seal the space between the flanges of wall tube and filter unit.
3. Mount the heating sleeve on the filter unit.
4. If applicable, install the compressed-air hoses between the filter unit and the back-purging unit.
into the sample gas inlet
8
Filter unit
5 Filter Unit
6 Probe tube
7 Flow direction
8 Sample gas inlet
of the filter unit.

Advertisement

Table of Contents
loading

Table of Contents