GAA610-M ADVANCED EMISSION GAS MONITORING SYSTEM | OI/GAA610-M-EN REV. B
... 13 Maintenance
... Replacing the filter element in the filter unit
9. Attach the new filter element and tighten it hand-tight with
the filter holding screw.
NOTICE
Damage to the filter element
Damage to the filter element due to improper installation.
Excessive force could break the filter element!
The Filter element might expand during the heating process.
Fasten the filter element only hand-tight.
1 Bridge
Abbildung 44: Insert the filter holder
10. Insert
A
the filter holder back into the Filter housing.
11. Turn the bridge counter-clockwise
2 Filter element
A
.
1 Filter housing
2 Filter holder
3 Filter retainer
Figure 45: Pusch in the filter holder
12. Turn the T-handle in clockwise direction.
This push the filter holder via the filter retainer in the
filter housing.
13. Press the filter into the housing until the two surfaces
aligned.
14. Turn the T-handle counterclockwise to remove the
mechanical force from the filter.
Turn the T-handle until there is a gap between the filter
holder and the filter retainer.
15. Close the cover of the filter unit.
16. Turn on power supply to the probe heater (–F63).
17. Deactivate the 'Maintenance Mode' on the control panel.
4 Brige
5 T-handle
A
are