GAA610-M ADVANCED EMISSION GAS MONITORING SYSTEM | OI/GAA610-M-EN REV. B
Factory setting
Passwords are pre-assigned for the factory-set user groups.
User Group
Access to password levels
Every user
0
Maintenance team
0, 1
Specialist team
0, 1, 2
Function block specialist
0, 1, 2, 3
Viewing Menu Items
All users can view all menu items, regardless of password level,
without entering a password.
Changing Menu Items
All users can make changes to password level 0 menu items
without entering a password.
Password level 1, 2 and 3 menu items can only be changed if the
user belongs to the group authorized for that level and after the
user's password has been entered.
Note
Entering the main menu and thus switching to the menu mode
can be password protected, refer to Access lock on page 46.
Change Privilege
After entering the password the user is authorized to change any
menu items on all password levels accessible at the user's level.
... 11 Operation
... Password protection
Duration of the change privilege
The change privilege remains in place until
•
either the gas analyzer automatically switches to
measuring mode if the user does not actuate a button for
about five minutes ('time-out'),
Password
•
or the user presses the 'MEAS' key twice in a row.
None
471100
If the user presses the 'MEAS' key only once to switch back to
081500
measuring mode, the change privilege initially remains in place.
325465
This is signaled by the flashing 'Password active' message
display.
In this way, the user does not need to enter the password again
before changing the menu items when switching back to menu
mode within the following approx. five minutes.
Note
The change privilege therefore refers to the time-limited
authorization to make changes to the menu items.
By contrast, the access right designates the principle
authorization defined by per configuration to make changes to
the menu items at certain password levels.
Change password
Refer to Chapter 'Configuration' in the operating instruction
OI/AO2000.