GAA610-M ADVANCED EMISSION GAS MONITORING SYSTEM | OI/GAA610-M-EN REV. B
... 2 Design and function
... System structure
Sample Gas Feed Unit SCC-F
1 Condensate monitor
2 LED display and reset switch for
condensate and flow alarm
Figure 4: Sample Gas Feed Unit SCC-F
The sample gas feed unit sucks the sample from the exhaust gas
and continuously feeds the gas analyzer with the sample gas.
The pump is a diaphragm pump. If the pump fails a flow alarm
occurs.
Flow control
The flow meters monitor the sample gas flow through the
analyzer.
The following parameters are considered:
•
Typical flow: 60 l/h
•
Minimum flow level: 40 l/h
•
Maximum flow level: 80 l/h
•
Flow error: 10 l/h
If the flow violates a limit value an alarm is displayed on the front
panel. In addition, the sample probe is automatically
backpurged.
3 Flow monitor with needle valve