GAA610-M ADVANCED EMISSION GAS MONITORING SYSTEM | OI/GAA610-M-EN REV. B
Analyzer cabinet – Visual inspection
Figure 35: Frontview of the analyzer system

Visual inspection

Pos.
Device, module
1
Main switch
2
Gas analyzer display: Measured values, flow
value, status messages, cabinet temperatur
3
Indicator light 'Common Alarm'
4
Indicator light 'Condensate Level Alarm'
5
Sample gas cooler temperature
6
Level of condensate collection bottle
7
Air filter of analyzer air conditioner
8
Sample gas flow
9
LEDs 'Condensate alarm' and 'Flow alarm'
0
Heated line temperature controller
k
Push button 'Measuring point'
l
Push button 'Backpurge Status'

Cleaning hints

Never use water or any solvents to clean parts inside the
analyzer cabinet.
Always operate the analyzer system with cabinet door closed.
Remove dust inside the analyzer cabinet using a broom and a
vacuum cleaner.
Clean the outside of the analyzer cabinet with a wet towel
and mild cleaning agents. Pay attention that no droplets
invade the cabinet.

Status LEDs

The three LEDs next to the screen show the user the gas
analyzer's status.
Status LED
Note
For detailed information on status messages and status signals
refer to Possible status messages on page 59.
Nominal condition
ON
Green LED 'Power' ON
OFF
OFF
3 °C
See page 83
See page 83
40 to 80 l/h
LEDs OFF
180 °C, ±1 °C
ON
OFF
Description
The green 'Power' LED lights when the power supply is on.
The yellow 'Maint' LED lights when the 'Maintenance request'
status signal is active.
The 'STATUS MESSAGE' softkey appears on the screen at the
same time.
The red 'Error' LED lights when the 'Failure' status signal or the
overall status signal is active.
The 'STATUS MESSAGE' softkey appears on the screen at the
same time.
73