Finding a Suitable Location for the
RamanMicro 300
The following conditions should be observed:
The environment must be suitable for a Class 3B laser system.
If your RamanMicro 300 has a manual stage there must be 1 power point for the
microscope in addition to the power points required for the spectrometer module. If
your RamanMicro 300 has a motorized stage there must be 2 power points (1 for the
microscope, 1 for the stage control box) in addition to the power points required for the
spectrometer module.
The workbench must be flat and level.
The workbench where the microscope is located should be at a suitable height to allow
the user to work comfortably.
Use conducting floor coverings to minimize static discharge.
Floor vibrations or noise from heavy manufacturing equipment can affect the
performance of the instrument.
Intense magnetic fields should be at least 5.5 meters (18 feet) away from the
spectrometer.
Never place the instrument near a window that gets direct sunlight.
Preparing Your Facility . 33