Summary - PerkinElmer RamanMicro 300 Series Getting Started Manual

Table of Contents

Advertisement

NOTE: The NIST intensity correction is valid in the recommended operating temperature
range 20 °C – 25 °C. If you move the instrument to an environment outside this
temperature range, that is 15 °C – 20 °C or 25 °C – 30 °C, we recommend that you
perform a NIST intensity correction at this new temperature.
Remember that if you have a fiber optic probe, you must intensity calibrate the fiber optic
probe as well as the microscope objective lenses. Ensure that the fiber optic probe has a lens
cap in place during intensity calibration of the objectives.

Summary

The only calibration procedure that needs to be performed on a regular basis is the
wavelength calibration verification procedure.
The wavelength and laser frequency calibrations should only be performed if the
instrument fails the wavelength calibration verification procedure.
The intensity calibration should only be performed on a yearly basis or if the
environment (temperature and / or humidity) changes significantly, or if the instrument
is moved or repaired.
Using the RamanMicro 300 . 87

Advertisement

Table of Contents
loading

Table of Contents