7. Select Include this location in the search, click Browse, and then navigate to
C:\Program Files\PerkinElmer\ServiceIR\Raman.
These options may be displayed automatically as you work through the wizard.
8. Click Next.
The Completing the Found New Hardware Wizard is displayed.
9. Click Finish.
The Found New Hardware Wizard closes.
Problem:
A motorized sample accessory does not go to the correct position
when asked.
Description
This means that for some reason the motorized stage is jammed. This
typically happens if the sample holder on the stage is different to that
selected on the Sampling tab in the Scan and Instrument Setup dialog.
Solution 1
Go to the Sampling tab and ensure that the selected sample holder is
correct. If it is not then select the appropriate sample holder.
Maintenance and Troubleshooting . 97