Table of Contents

Advertisement

3. Select "Elevation".
• The tilt screen is displayed if the
instrument is out of level.
Level the instrument.
"7.2 Levelling"
4. Sight the first known point and
press [MEAS] to begin
measurement. Press [STOP].
The measurement results are
displayed on the screen.
5. Press [YES] to use the
measurement results of the first
known point.
6. Input the known point. After
setting the elevation for the first
known point, press [NEXT] to
move to the second point.
7. If measuring two or more known
points, repeat procedures 4 to 6 in
the same way from the second
point.
• Press {ESC} to return to the
previous known point.
8. Press [CALC] to automatically
start calculations after
observations of all known points
are completed. Instrument station
elevation and standard deviation,
which describes the measurement
accuracy, are displayed.
13. SETTING INSTRUMENT STATION
Resection.
NEZ
Elevation
Resection 1st PT
SD
ZA
8 0
HA-R
1 4 0
1 s t P T
Z p :
H R
LOAD
REC
4 2
1 5
4 9
1 5
MEAS
1 1 . 8 9 1
0 . 1 0 0 m
NEXT
CALC