Restarting the MS System After a Shutdown

Restarting the Data System

Restarting the Data System Using Windows

Restarting the Data System Using the Power Button

Thermo Scientific
When you restart the MS system after a complete shutdown, the sequence of tasks is
important.
• Restarting the Data System
• Resetting the Mass Spectrometer
• Pumping the MS Down
• Resetting Calibration Parameters
• Starting the LC, Gases, and Autosampler
If possible, use the Windows 7 or Windows 10 restart procedure the data system so that
Windows can properly close applications and save changes to any open application.
Note After you reset the data system, the communications link between the data system
and the MS is automatically reestablished. When this occurs, the Communication LED
turns yellow and then green. If the system is unable to reestablish the communications
link, follow
"Resetting the Mass Spectrometer."
1. On the Windows taskbar, choose Start, and then click the arrow next to Shut Down.
2. Choose Restart, and then click OK.
1. Press the power button on the data system computer.
2. Wait at least 20 seconds.
3. Press the power button again.