Reset; Resetting The System Settings - Xylem WTW Turb 430 IR/T Operating Manual

Handheld turbidimeter
Hide thumbs Also See for WTW Turb 430 IR/T:
Table of Contents

Advertisement

Turb 430 IR/T
ba75507d09
08/2021
The data are output in the following order:
1
1
2
2
15.01.21;15:06:49;1;26.5;NTU;VALID;[3-P StdCAL];15.01.21;08:67:47
15.01.21;16:06:49;1;16.5;NTU;VALID;[3-P StdCAL];15.01.21;08:67:47
1 Date of storing
2 Time of storing
3 Selected ID
4 Measured value or
Upper/lower measuring range limit
(only with measured value status, OFL/UFL)
5 Unit of the measured value
6 Measured value status
* VALID: Measured value valid
* INVALID: Measured value invalid
* UFL: Measured value below the lower measuring range limit
* OFL: Measured value above the upper measuring range limit
7 Calibration status: Calibration type that was selected for the
measurement
8 Calibration status: Date of calibration
9 Calibration status: Time of calibration
If the connected external printer does not print, please check
whether the same baud rate is set at the Turb 430 IR/T and print-
er.

4.10 Reset

You can reset (initialize) all system and measurement settings.

4.10.1 Resetting the system settings

With the System / Rücksetzen function, all resettable settings are reset.
 Settings for Trübung (see section 4.10.2)
 System settings
System setting
Sprache
Baudrate
3
3
4
4
5
5
6
6
Default settings
English
4800 Baud
7
7
8
8
Operation
9
9
41