Reset; Resetting The Measurement Settings - Xylem SI Analytics HandyLab 200 Operating Manual

Conductivity meter
Table of Contents

Advertisement

HandyLab 200
ba77063e03
01/2018
4.5.2 D
ISPLAYING AND EDITING THE MEASUREMENT DATA MEMORY
4.6

Reset

You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.

4.6.1 Resetting the measurement settings

Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
The following settings for conductivity measurements are reset to the
default settings with the Reset function:
Setting
Cal. interval
Measured parameter
Cell constant (c)
Temperature compensation
Reference temperature
Temperature coefficient (TC) of
the linear temperature compen-
sation
TDS factor
Stability control
Temperature unit
The sensor settings are reset under the Reset menu item in the menu
for calibration and measurement settings. To open the settings, acti-
vate the relevant measuring window in the measured value display and
press the <F1>/[Menu] key shortly.
Default settings
150 d
ϰ
calibrated: 0.475 cm
-1
adjusted: 0.475 cm
-1
calibrated: cm
adjusted:
nLF
25 °C
2.000 %/K
1.00
On
°C
Operation
.
-1
35

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SI Analytics HandyLab 200 and is the answer not in the manual?

Questions and answers

Table of Contents