Reset; Resetting The Measurement Settings - Xylem YSI MultiLab 4010P-1 Operation Manual

Table of Contents

Advertisement

Settings
pH
70
You can also generally switch on the display illumination (see section 9.5.1
S
, page 68).
YSTEM
9.6

Reset

You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
9.6.1

Resetting the measurement settings

The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
The following settings for pH measurements are reset to the default settings
with the Reset function:
Setting
Buffer
Calibration interval
Unit for slope
Measured parameter
Unit for slope
mV resolution
Asymmetry
Slope
Man. temperature
Single-point calibration
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <MENU/
ENTER> key.
Default settings
YSI
7 d
mV/pH
pH
0.001
0.1
0 mV
-59.2 mV
25 °C
off
MultiLab 4010-1