Cleaning; Cleaning The Cell Shaft; Cleaning The Cells - Xylem WTW Turb 430 IR/T Operating Manual

Handheld turbidimeter
Hide thumbs Also See for WTW Turb 430 IR/T:
Table of Contents

Advertisement

Maintenance, cleaning, disposal
46
5.2

Cleaning

Occasionally wipe the outside of the measuring instrument with a damp, lint-
free cloth. Disinfect the housing with isopropanol as required.
NOTE
The housing components are made of synthetic materials (polyure-
thane, ABS and PMMA). Thus, avoid contact with acetone and similar
detergents that contain solvents. Remove any splashes immediately.

5.2.1 Cleaning the cell shaft

If liquid is in the cell shaft (e.g. due to a spilled cell), clean the cell shaft as
follows:
1
Switch the Turb 430 IR/T off and pull out the power plug.
2
Rinse the cell shaft with distilled water.

5.2.2 Cleaning the cells

Cells have to be clean, dry, and free of fingerprints. Therefore, clean them
regularly:
1
Clean the cells inside and out with hydrochloric acid or laboratory
soap.
2
Rinse several times with distilled water.
3
Let them dry in the air.
4
Only hold the cells by the top or by the light protection cap so that the
optical path is not impaired.
5
Before measuring, clean the cell with the enclosed cleaning cloth.
Turb 430 IR/T
ba75507d09
08/2021