Reset; Resetting The Calibration Values; Resetting The Measurement And System Settings - Xylem SI Analytics Lab 855 Operating Manual

Ph meter
Hide thumbs Also See for SI Analytics Lab 855:
Table of Contents

Advertisement

Reset

8
Reset
32
You can erase the calibration values and reset (initialize) the measurement and
system settings.
8.1

Resetting the calibration values

1.
Press <On/Off__> to open the menu for the reset of the calibration data.
InIt.C is displayed.
2.
Use <><> to display no or YES.
YES: Reset the calibration values
no: Retain the calibration values.
3.
Confirm with <ENTER>.
The menu is finished. The meter switches to the measuring mode.
The calibration values are reset to default. All other meter
settings are retained.
Recalibrate after performing a reset.
Calibration values to be reset
Calibration value
Zero point (ASY )
Slope (SLO )
8.2

Resetting the measurement and system settings

1.
Switch on the meter with <On/Off>.
The display test appears briefly on the display.
2.
During the display test, press <M> to open the menu for the reset of the
meter settings.
Init is displayed.
3.
Use <><> to display no or YES.
YES: Reset the meter settings.
no: Retain the meter settings.
4.
Confirm with <ENTER>.
The settings are reset. The menu is finished.
The meter switches to the measuring mode.
Default settings
0 mV (pH 7.000)
-59.16 mV/pH (100 %)
ba77143e02
Lab 855
12/2017