Checking the function blocks and displays
1 Checking the "deadman" switch function
1.1 Exposure
An exposure value is set at the control console.
- Activate the exposure with the exposure switch and end it immediately by releasing the
switch.
The exposure must end immediately.
1.2 Fluoroscopy
Activate fluoroscopy for a short time and release the fluoroscopy switch.
Fluoroscopy must stop immediately.
a) with the fluoroscopic switch at the control console
b) with the foot switch
Deadman's switch is functioning
2 Radiation indicators
Check that the radiation indicators at the control console and on the monitor cart are visi-
ble during radiation and that an audible (whistling) signal is sounded after exposure
release for about 1.5 seconds.
Radiation indicators are functioning
Siemens AG
Medical Solutions
SPR2-310.662.01
Rev. 02
05.06
4
YES
NO
YES
NO
Page 1 of 2
CS PS SP
4 - 1
4
4
4
4
ARCADIS Varic