:ERRor?
:VERSion?
:SECurity
:POBuffer <numeric value>
GPIB Address Change
:SYSTem:COMMunicate:GPIB:ADDRess
This command is used to set the GPIB address.
Arguments
Type:
Range:
Rounding:
Command Type:
Setting
Syntax:
Example:
Query
Syntax:
Response:
Considerations:
1) Setting the address to 31 puts the instrument in the off-bus state.
2) Using the MAX option sets the address to 30, not 31.
Default Power-on is address 9.
Error Queue Reading
:SYSTem:ERRor?
This query returns the first entry in the error queue, and removes that entry from the queue. It's
function is identical to that of the :STATus:QUEue:NEXT? query.
Command Type:
Query
Syntax:
Response:
SCPI Version
:SYSTem:VERSion?
1.800.544.2843
:GPIB
:ADDRess <numeric value>
[:STATe] <Boolean>
Numeric
0 to 31
to integer value
Setting or Query
:SYSTem:COMMunicate:GPIB:ADDRess<ws><address>|MINimum|MAXimum
:SYST:COMM:GPIB:ADDR 20
:SYSTem:COMMunicate:GPIB:ADDRess?[<ws>MINimum|MAXimum]
<address> in NR1 format
Query only
:SYSTem:ERRor?
<Error number>, "<error description>"