B+K precision 4075B User Manual page 68

Arbitrary/function waveform generator
Table of Contents

Advertisement

Examples:
Response:
Considerations:
1) The MAXimum amplitude is dependent on the offset.
2) The MAX and MIN arguments should not be used in a program message containing an
Offset
:SOURce:VOLTage:OFFSet <offset>
The offset command is used to set the DC offset of the output waveform. Note that the amplitude and
the offset are limited by the relation
Peak Amplitude + |Offset| ≤ 5V
Arguments
Type:
Units:
Range:
Rounding:
Command Type:
Setting
Syntax:
Examples:
Query
Syntax:
Examples:
Response:
Considerations:
1) The MAXimum offset is dependent on the amplitude.
2) The MAX and MIN arguments should not be used in a program message containing an
1.800.544.2843
[<ws>MINimum|MAXimum]
:VOLT:AMPL?
:VOLT:AMPL? MAX
NR2
OFFSet command, since these values are evaluated during parsing, based on the current
value of the offset.
Numeric
V, mV
10mV to 4.99V
to 10mV
Setting or Query
[:SOURce]:VOLTage:OFFSet<ws><offset>[units]
[:SOURce]:VOLTage:OFFSet<ws>MINimum|MAXimum
:VOLT:OFFS 2.5
:VOLT:OFFS 2.5V
:VOLT:OFFS MAX
[:SOURce]:VOLTage:OFFSet?[<ws>MINimum|MAXimum]
:VOLT:OFFS?
:VOLT:OFFS? MAX
NR2
AMPLitude command, since these values are evaluated during parsing, based on the current
value of the amplitude.