B+K precision 4075B User Manual page 63

Arbitrary/function waveform generator
Table of Contents

Advertisement

Type:
Syntax:
Examples:
Query
Syntax:
Response:
c)
*ESR? - Standard event status register query
This query is used to read the value of the Standard Event Status Register. Reading the register
clears it.
Type:
Syntax:
Response:
d)
*PSC - Power on status clear command
This command is used to control the automatic power-on clearing of certain status functions.
Arguments
Type:
Type:
Command
Syntax:
Examples:
Query
Syntax:
Response:
When set to ON (1), the Service Request Enable Register and the Standard Event Status Enable
Register are cleared on power-on.
e)
*SRE - Service request enable command
This command sets the Service Request Enable Register bits.
Arguments
Type:
Range:
Type:
Syntax:
Examples:
Query
1.800.544.2843
Common Command or Query
*ESE<ws><NRf>
*ESE 48 (Enables the CME and EXE bits)
*ESE 255 (Enables all standard events)
*ESE?
<NR1>
Common Query
*ESR?
<NR1>
Boolean
Common Command or Query
*PSC<ws><Boolean>
*PSC ON or *PSC 1
*PSC OFF or *PSC 0
*PSC?
ASCII 0 for OFF
ASCII 1 for ON
NRf
0 to 255. Non integer arguments are rounded before execution.
The value of bit 6 is ignored, and is set always to zero.
Common Command or Query
*SRE<ws><NRf>
*SRE 48 (Enables reporting of ESB and MAV events)
www.calcert.com
15
10
20
5
25
0
30
50
sales@calcert.com