1) The MIN | MAX arguments should be used only in a Program Message that does NOT
2) The MIN and MAX arguments refer to currently settable minimum or maximum.
3) FIXed is alias for CW.
Amplitude
:SOURce:VOLTage[:AMPLitude] <p-p amplitude>
The amplitude command is used to set the peak-to-peak amplitude of the output waveform. Note that
the amplitude and the offset are limited by the relation
Peak Amplitude + |Offset| ≤ 5V
Arguments
Type:
Units:
Range:
Rounding:
Command Type:
Setting
Syntax:
Examples:
Query
Syntax:
1.800.544.2843
Fmin = 1/(100 S * Wavelength)
The value is rounded to 4 digits.
Setting or Query
[:SOURce]:FREQuency[:CW]<ws><frequency>[units]
[:SOURce]:FREQuency<ws>MINimum|MAXimum
:FREQ 5KHZ
:FREQ 5E3
:FREQ MAXIMUM
:FREQ MIN
[:SOURce]:FREQuency[:CW]?[<ws>MAXimum|MINimum]
:FREQ?
:FREQ? MAX
NR3
contain Program Message Units specifying Arbitrary Point Rate or Wavelength, since the
MAXimum or MINimum value is calculated at the time the command is parsed.
Numeric
V, mV, VPP, mVPP
10mV to 10V
1mV from 10mV to 999mV. 10mV from 1V to 10V.
Setting or Query
[:SOURce]:VOLTage:AMPLitude<ws><amplitude>[units]
[:SOURce]:VOLTage:AMPLitude<ws>MINimum|MAXimum
:VOLT:AMPL 2.5
:VOLT:AMPL 2.5V
:VOLT:AMPL MAX
[:SOURce]:VOLTage:AMPLitude?
www.calcert.com
15
10
20
5
25
0
30
54