Example
To determine the correct weight of a sample, enter the density of this sample. In this example, use the air density saved in the balance.
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 1: Air buoyancy correction: Air buoyancy determination: On
Step
1. Turn on the balance and
configure as indicated above
2. Start air buoyancy correction
3. Enter the density of your sample
(in this example: 1.0120 g/cm
4. Save the density of the sample
as Rho
; this simultaneously activates
W
air buoyancy correction
5. Place the sample on the balance
(for example, 42.12648 g)
6. Display the saved air density,
if desired
7. Exit readout of air density
90
Press key(s) (or follow instructions)
e
Start
soft key
1.01
20
3
)
RhoW
soft key
Load balance
RhoA
soft key
oo
soft key
Display/Output
Max 210 g
0%
F
4212648
S
CORR.: RhoW = 1.0120000 g/cm3
Cal
PT1/T1
RhoA Weigh. Fact.
Max 210 g
0%
F
4212648
S
AIR D: RhoA = 1.2000000 kg/m3 (Def)
PT1/T1Param.
oo
d=0.01mg
100%
G
L
a
g
S
d
T
S-ID
d=0.01mg
100%
G
L
g
S
d
T
Wvgen
Start
S-ID