Fuji Electric ZPB Insruction Manual page 70

Ndir type infrared gas analyzer
Hide thumbs Also See for ZPB:
Table of Contents

Advertisement

ENT
key.
memorized.
Note: This setting time is very impor-
tant setting when sample switch
is carried out.
Do not change the setting time
unless it is required.
Explanation
Set up switching time and measuring time of the reference gas.
Switching time of reference gas : 1 to 30s (Initial value 5 sec)
Measuring time of reference gas : 1 to 60s (Initial value 5 sec)
Note) Switching time of the sample gas, measuring time of the sample gas, flow time of the
reference gas and flow time of the sample gas
are updated automatically based on following relational expression.
Switching time of sample gas (t1) = Switching time of reference gas (t3)
Measuring time of sample gas (t2) = Measuring time of reference gas (t4)
Flow time of sample gas (Ts) = Flow time of reference gas (Tr)
Flow time of reference gas (Tr) =
Switching time of reference gas (t3) + measuring time of reference gas (t4)
Flow time of sample gas (Ts) =
Switching time of sample gas (t1) + measuring time of sample gas (t2)
1 cycle = flow time of reference gas (Tr) + flow time of sample gas (Ts)
-
ENT
60
Maintenance
Set the gas
Mode
replacement time.
Sample Switch
1 to 30 sec
Gas Flow Time
REF. Gas:
Replacement Time
REF. Gas:
Measuring Time
REF. Gas:
Flowing Time
SMP. Gas:
Replacement Time
SMP. Gas:
Measuring Time
SMP. Gas:
Flowing Time
Maintenance
Select operating item
Mode
Sample Switch
Gas Flow Time
REF. Gas:
Replacement Time
REF. Gas:
Measuring Time
REF. Gas:
Flowing Time
SMP. Gas:
Replacement Time
SMP. Gas:
Measuring Time
SMP. Gas:
Flowing Time
05
sec
05
sec
10
sec
05
sec
05
sec
10
sec
ENT
05
sec
05
sec
10
sec
05
sec
05
sec
10
sec
INZ-TN2ZPB-E