Fuji Electric ZPB Insruction Manual page 88

Ndir type infrared gas analyzer
Hide thumbs Also See for ZPB:
Table of Contents

Advertisement

21-On
C =
21-Os
correction
C
S
gas
O
S
O
n
(value changeable by setting.0 to
Average value after O
correction and O
2
calculation:
The result of O
taneous O
average value over the preset period of
time.
Moving average method is used. Sampling
interval is 30 secs.
(Output is updated every 30 seconds.
Update is the averaged value of the most
recently elapsed averaging time period.)
to 59 minutes (in increments of 1 minute)
or 1 to 4 hours (in increments of 1 hour).
Average value resetting:
The above-mentioned output of average
value is started from the initial state by
opening the average value resetting input
terminals after short circuiting for 1.5 sec
or longer.
Output is reset by input voltage and
restarted by opening the terminal circuit.
Communication function:
RS-485 (9pins D-sub connector)
Half-duplex bit serial
Start-stop synchronization
ModbusTM protocol
Atmospheric pressure correction:
Measure atmospheric pressure and cal-
culate compensation (for use, be sure to
relieve the exhaust gas from analyzer to
the atmosphere)
× Cs
concentration (Limit
2
)
2
concentration
2
)
2
average value
2
correction or instan-
2
value can be output as an
2
Read measurement con-
centration and instrument
status.
232C interface, an RS-232C
RS-485 converter should
be used.
sured value or less relating
to the change of the atmo-
4. Performance
Repeatability:
Linearity:
prior to atmospheric pressure correction
2
(option)
Zero drift:
component of NDIR)
Span drift:
Response time (T
) :
90
30 seconds or better
Response interval may be changed de-
pending on timing of the switching gas
by sample switching operation. (Td=5-20
seconds)
Interference from other gases:
Sample switching design effectively
minimizes interference. But it may occur
depending on component gas and its
concentration.
ence in this case.
Contact manufacturer for application
5. EC Directive Compliance
LVD (2014/35/EU)
EN 61010-1
EN 62311
EMC (2014/30/EU)
EN 61326-1 (Table 2)
EN 55011 (Group 1 Class A)
EN 61000-3-2 (Class A)
EN 61000-3-3
EN 61326-2-3
RoHS (2011/65/EU)
EN 50581
6. Requirements for Sample Gas
Flow rate:
1.0L / min ±0.2L / min
Temperature:
Pressure:
10 kPa or less (Gas outlet side should
be open to the atmospheric air.)
Dust:
100 μg/Nm
μm or smaller
Mist:
Unallowable
Moisture:
Less than 2°C saturation point. (Contain
comparable sample gas and reference
gas)
Corrosive component:
1 ppm or less
-
78
sensor)
2
or less in particle size of 0.3
3
INZ-TN2ZPB-E
-

Advertisement

Table of Contents
loading

Table of Contents